Thermo Scientific™

PXS4-611W 60kV X-Ray Sources

Catalog number: PXS4611W
Thermo Scientific™

PXS4-611W 60kV X-Ray Sources

Catalog number: PXS4611W
Thermo Scientific™ PXS4-611W, 60kV X-ray Source is a constant potential, continuous duty X-ray source for use in field applications or where high stability is required. Side-window sealed tube and a high voltage power supply.
 
Catalog Number
PXS4611W
Unit Size
Each
Cone of Illumination
35° inclusive
Source
Focused filament
Power Dissipation
60w
Price (USD)
Full specifications
Cone of Illumination35° inclusive
Window MaterialBeryllium
DescriptionX-ray source 60kV
Diameter (English) Window0.37 in.
Diameter (Metric) Window9.4mm
Max. Beam Current1mA
Operating Voltage Range5 to 60kV
Target MaterialTungsten, Copper (Standard)
Temperature (Metric) Operating0° to 40°C (altitude of 5,000 ft.)
Thickness (English) Window0.005 in.
Thickness (Metric) Window0.127mm
Voltage12 ± 2VDC
Weight (English)9 lb.
Weight (Metric)4.1kg
Item DescriptionX-ray source 60kV
Power Dissipation60w
SourceFocused filament
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizeCone of IlluminationSourcePower DissipationPrice (USD)
PXS4611WFull specifications
Each35° inclusiveFocused filament60wRequest A Quote
Cone of Illumination35° inclusive
Window MaterialBeryllium
DescriptionX-ray source 60kV
Diameter (English) Window0.37 in.
Diameter (Metric) Window9.4mm
Max. Beam Current1mA
Operating Voltage Range5 to 60kV
Target MaterialTungsten, Copper (Standard)
Temperature (Metric) Operating0° to 40°C (altitude of 5,000 ft.)
Thickness (English) Window0.005 in.
Thickness (Metric) Window0.127mm
Voltage12 ± 2VDC
Weight (English)9 lb.
Weight (Metric)4.1kg
Item DescriptionX-ray source 60kV
Power Dissipation60w
SourceFocused filament
Unit SizeEach
Showing 1 of 1
  • Highly stable flux output of <0.2% standard deviation over 4hr period
  • 15.8mm Focus-to-Object Distance (FOD)
  • Water-cooled, air cooled options (PXS4-612CU and PXS4-613WH) available
  • Target voltage and electron beam current adjusted using a separate external variable DC voltage source
  • Designed with five mounting holes, four in the baseplate and one near rear connector
  • Window orientation relative to housing is factory set to position C
  • Shielding of housing is sufficient to ensure that X-ray leakage is <0.5 mR/hour

Recommended for:

  • Gauging